| Display | LCD: LED backlight | ||
|---|---|---|---|
| Environment | Installation location | Indoors, at a height of up to 2000 m | |
| Spec guaranteed range temperature/ humidity | 5 °C to 35 °C (41 °F to 95 °F)/20 %rh to 80 %rh (no condensation) | ||
| Operating range temperature/ humidity | 0 °C to 40 °C (32 °F to 104 °F)/20 %rh to 80 %rh (no condensation) | ||
| Storage range temperature/ humidity | -20 °C to 70 °C (-4 °F to 158 °F)/90 %rh or less (no condensation) | ||
| Power supply | Nominal voltage range (allowable voltage range) | 100 VAC to 240 VAC (90 VAC to 250 VAC) | |
| Power consumption | When no load is connected (READY) | 100 VA or less | |
| When rated load isconnected | 800 VA max. | ||
| Allowable frequency range | 47 Hz to 63 Hz | ||
| Insulation resistance (between AC LINE and the chassis) | 30 MΩ or more (500 VDC) | ||
| Withstanding voltage (between AC LINE and the chassis) | 1500 VAC, one minute | ||
| Earth continuity | 25 AAC, 0.1 Ω or less | ||
| Electromagnetic compatibility (EMC) *1 | Complies with the requirements of the following directive and standard. EMC Directive 2014/30/EU EN 61326-1(ClassA *2), EN 55011(ClassA *2, Group1 *3) EN 61000-3-2, EN 61000-3-3 Applicable under the following conditions The maximum length of all cabling and wiring connected to the TOS5200 must be less than 2.5 m. The shielded cable is being used when using the SIGNAL I/O. The high test lead TL31-TOS | ||
| Safety *1 | Complies with the requirements of the following directive and standard. Low Voltage Directive 2014/35/EU, EN 61010-1 (Class I *4 Pollution degree 2) | ||
| Dimensions (mm(inches))(maximum) | 320 (12.6″) (330(12.99″)) W × 132(5.2″) (150(5.91″)) H × 350(13.78″) (420(16.54″)) D | ||
| Weight | Approx. 14 kg (30.9 lbs) | ||
| Accessories | Power cord : 1pc. / High test lead (TL31-TOS) : 1set (1 red wire and 1 black wire, each with alligator clips); 1.5 m / D-sub 25-pin plug : 1set ; assembly type / High-voltage warning sticker : 1pc. / Setup Guide / Quick Reference(1 each for English and Japanese) / Safety informaion / CD-R *5 | ||
*1: Only on models that have the CE marking on the panel. Although signals are insulated with output terminals, each signal is common. Logic setting is also possible. *2: This is a Class A equipment. This product is intended for use in an industrial environment. This product may cause interference if used in residential areas. Such use must be avoided unless the user takes special measures to reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts. *3: This is a Group 1 equipment. This product does not generate and/or use intentionally radio- frequency energy, in the form of electromagnetic radiation, inductive and/or capacitive coupling, for the treatment of material or inspection/analysis purpose. *4: This is a Class I equipment. Be sure to ground this product’s protective conductor terminal. The safety of this product is only guaranteed when the product is properly grounded. *5: Contains the User’s Manual, the Cimmunication Interface Manual, VISA library (KI-VISA), IVI-COM driver, and Safety evaluation test.